Reusable monitor nibp cuff single&double hose
Intended use
Collect and transmit human blood pressure signals through vasoconstriction and expansion of the pressure on the inner bladder of the cuff. For non-invasive blood pressure measurement of patients.



Product classification
This product have large adult, adult, small adult, child, infant and neonatal cuffs according to the applicable objects. According to the service life, have reusable NIBP cuff and disposable NIBP cuff; according to the classification of rubber tube,have Single and double tube NIBP cuffs
Basic performance
1. Maximum inflation pressure:
① Neonatal and infant cuff: the maximum inflation should reach 180mmHg (24kPa);
② Child, small adult, adult, large adult, and thigh cuff: the maximum inflation can reach 360mmHg (48kPa)
2. Air tightness: should not exceed 4mmHg/min.
3. Durability:
① reusable cuff opening and closing cycles ≥ 1000 times, inflation and deflation ≥ 10,000 times.
② disposable cuff opening and closing cycles ≥ 500 times, inflation and deflation ≥ 5000 times.